Rhythm Biosciences investigators reported development of ColoSTAT, a multiplexed ELISA that quantifies five serum protein biomarkers in a single assay intended to support triage for symptomatic patients being considered for colonoscopy. The authors combined samples from two prospective cross-sectional studies with retrospective biobanked serum to assemble a training cohort of 300 samples and a validation cohort of 200 samples matched by sex, age and clinical diagnosis. The validation set was then supplemented with 100 additional serum samples from patients with advanced adenoma (AA). The overall sample set in the report had a median age of 64 years (range 28–91) and 56% male representation. The source did not report the individual biomarker names in this summary; those analytic details were not provided in the preprint text excerpt.
When compared against colonoscopy as the reference standard, the authors reported that ColoSTAT detected colorectal cancer across all stages with an overall sensitivity of 89% and a specificity of 54%. For the age subgroup of 45 to 84 years the test sensitivity and specificity were 90.3% and 54.3%, respectively, according to the study team. The manuscript modelled test operating characteristics at an intended-use CRC prevalence of 4%; in that scenario the negative predictive value (NPV) exceeded 99% and the positive predictive value (PPV) exceeded 7% as reported by the authors. The paper noted that these predictive values were modelled rather than observed in a screened primary-care population.
The investigators provided stage-stratified sensitivity estimates. Early-stage CRC sensitivity was 90.7%; late-stage sensitivity was 89.4%. The team also assessed detection of advanced adenoma, reporting 58% sensitivity in the supplemented validation set. The authors framed these numbers as indicating the test’s potential to identify clinically significant neoplasia that would merit prioritised diagnostic colonoscopy. The source text did not report separate specificity figures for advanced adenoma or provide lesion-size or location breakdowns in the excerpt available.
Using an assumed CRC prevalence of 4% to reflect an intended-use symptomatic population, the study team modelled a very high NPV (>99%) and a modest PPV (>7%). Those modelling outputs imply the assay could function primarily as a rule-out tool to reduce unnecessary urgent colonoscopies among low-risk symptomatic patients. Age-stratified performance was presented for the 45–84-year group, yielding slightly higher sensitivity (90.3%) versus the overall cohort, while specificity remained near 54%. The source did not provide separate analyses by sex, comorbidity, or specific presenting symptoms in the accessible text.
The report used a case-control design enriched for cancer cases. The authors acknowledged that the cohorts were assembled from prospective cross-sectional studies and retrospective biobanks, and that enrichment for cancer cases can affect apparent sensitivity and specificity compared with a purely prospective primary-care triage population. All authors were current or former employees of Rhythm Biosciences Limited; four named authors also hold company stock, a financial relationship disclosed in the manuscript. Ethical approval was granted by the Central Adelaide Local Health Network Human Research Ethics Committee, and the authors stated that required participant consents had been obtained.
The manuscript indicates that the study protocol and datasets generated during the work will be made available on reasonable request to the authors. The preprint does not report regulatory approvals, intended clinical pathways, or prospective implementation data in a real-world primary-care setting. That absence leaves unanswered questions about how ColoSTAT would perform in an unselected symptomatic population, how it would integrate into referral pathways, and what clinical threshold would be adopted to balance missed cancers against colonoscopy demand.
Limitations documented in the source include the case-control enrichment and the use of modelled predictive values rather than observed PPV and NPV in a representative symptomatic cohort. The authors did not report full analytic details of the five biomarkers or the assay cutoffs in the provided abstract text; those specifics would be necessary to evaluate reproducibility, laboratory requirements, and potential external validation. Data access is available on reasonable request, which may permit independent validation but does not substitute for prospective outcome data in routine clinical practice.