Investigators reported strong agreement between different plasma assays when measuring biomarker levels at single time points, but much weaker agreement when estimating individual rates of change. In the ADNI cohort (393 participants, median follow-up 5.0 years), multiple assays assayed aliquots of the same samples and returned similar level estimates: intraclass correlations were 0.75–0.95 for pTau217, pTau181, GFAP, and NfL. Short, clear fact: single-time-point measurements align. Yet the slope story differed. The cohort showed substantially lower concordance for individual slopes — for example, the ICC for pTau217 change across assays was 0.44 (95% CI 0.37–0.50). The study team observed that individual longitudinal trajectories were dominated by random disagreement rather than consistent assay-specific bias.
The study analysed 15 measures of five plasma biomarkers on aliquots of the same samples to separate shared longitudinal change from assay-specific variation. Investigators then tested their approach externally in 1,093 participants from the A4/LEARN datasets. The authors noted the analysis used three samples per person in ADNI to estimate individual slopes; they did not report experimental assay protocols beyond noting multiple assay measures and that aliquots were identical for cross-assay comparison. The report is a preprint and has not been peer reviewed.
Data indicated that assay-specific slope variance was approximately zero, which implies that disagreement in longitudinal change was mainly random noise rather than systematic differences between assays. Single-assay reliability for three-sample slopes was low across biomarkers: the reported reliabilities were at most 0.40 (0.34–0.46) for pTau217, 0.42 (0.33–0.51) for pTau181, 0.47 (0.37–0.60) for GFAP, 0.52 (0.40–0.64) for NfL, and 0.10 (0.03–0.20) for Aβ42/40. The authors calculated that two pTau217 measurements would need to differ by 65%–131% (assay-dependent) to exceed measurement noise — a stark demonstration of how within-person variability can overwhelm expected biological change.
The investigators modelled sample-size consequences of low slope reliability. Detecting a 25% slowing of pTau217 longitudinal change required very large participant numbers per arm under their assumptions: reported ranges extended from hundreds to several thousands depending on assay performance and whether measures were averaged. Averaging replicated endpoint measures reduced sample-size requirements substantially in their models (examples in the report show reductions from thousands to mid-hundreds per arm). Short statement: low reliability inflates trial size and cost. The study team observed that endpoint replication or averaging, and selecting higher-reliability assays, materially lowers trial sample-size demands.
In the external A4/LEARN dataset (1,093 participants), single-assay pTau217 reliability was reported at 0.54 (0.46–0.61). Extending the observation span produced measurable gains: increasing the span reported from 4.8 to 7.3 years improved reliability by 0.10 (0.00–0.22). The authors noted that adding a fourth intermediate sample within that longer span added little to reliability, suggesting that total follow-up duration matters more than the number of intermediate draws. The cohort analysis therefore favoured longer intervals between first and last sample over denser sampling for improving slope estimates.
Authors concluded that with three single-assay samples much of an individual's measured change is noise. They recommended prioritising longer observational spans, endpoint replication (averaging or repeated endpoints), and careful assay selection for serial plasma testing and for trials where plasma slopes are an outcome. The report also communicated trial design consequences quantitatively and validated key findings in an independent dataset.
Limitations cited in the source: this work is a secondary analysis of de-identified data from completed studies and is reported as a medRxiv preprint not yet peer reviewed. The authors did not report full assay procedural details in the abstract; specific lab protocols, calibration procedures, or assay manufacturers were not described in the text provided. Remaining clinical questions documented by the authors include which specific assay platforms and analytic strategies will most efficiently reduce random slope noise, and the optimal trade-offs between follow-up duration and the number of samples for therapeutic trials. These items remain to be resolved in peer-reviewed, prospectively designed studies.