The study team recorded pelvic floor responses during sacral neuromodulation (SNM) lead implantation in nine participants. During surgery, needle electrodes were placed to capture external anal sphincter (EAS) electromyography (EMG). A high-resolution manometry catheter obtained multi-segment urethra pressure measurements. Investigators performed intra-operative sacral nerve stimulation (SNS) while simultaneously recording EMG and pressure signals. The authors compared the SNS-evoked pressure responses to a previously collected dataset of pudendal nerve stimulation (PNS) responses to examine differences in muscle recruitment between the two nerve targets.
The report lists the recording modalities and the intra-operative stimulation paradigm in summary form. The source did not provide step-by-step parameter tables for all stimulation waveforms or a full breakdown of stimulation durations in the abstract; those methodological details may be present in the full text or supplementary materials but are not reported in the abstract excerpt.
Investigators quantified thresholds in relation to the visual motor threshold (VMT) identified intra-operatively. The applied current threshold required to evoke a measurable urethral response was significantly higher than VMT (p = 0.04), the authors reported. The threshold to activate the EAS was also higher than VMT but did not achieve statistical significance (p = 0.06). The team therefore observed that functional activation of the urethra required more current than the visually identified motor response, while EAS activation only trended toward higher thresholds. These observations were drawn from the nine participant recordings collected during SNM lead implantation.
The cohort showed similar maximum absolute urethral pressure changes when comparing SNS-evoked responses to previously collected PNS-evoked responses; the difference did not reach significance (p = 0.09). That finding indicates that, in absolute terms, the peak urethral pressure changes produced by stimulating either nerve could be comparable. The authors emphasized, however, that comparable peaks do not imply equivalent efficiency of stimulation.
Data indicated differences in efficiency between the two stimulation targets. Current amplitudes required to elicit the maximum change in urethral pressure were lower for SNS than for PNS (p = 0.01), the investigators reported. In addition, the maximum change in urethral pressure per milliampere of applied stimulation was higher for SNS than for PNS (p = 0.005). Put simply, SNS produced greater pelvic floor recruitment for each unit of current delivered in this comparison. The authors framed these results as evidence that sacral stimulation is more efficient at recruiting urethral responses than pudendal stimulation when measured as response per mA.
The study team suggested a mechanistic rationale rooted in electrode–nerve proximity. Typical SNM leads are positioned closer to the sacral nerve target than leads used off-label for pudendal stimulation, the authors noted. This closer apposition could reduce the current required to activate target fibers and thereby increase recruitment per mA. The source text links the observed efficiency differences to lead position as a plausible explanatory factor rather than to intrinsic superiority of a stimulation waveform; that mechanistic interpretation derives from anatomical and device-placement considerations described by the authors.
The authors explicitly describe this work as the first human comparison of muscle recruitment between sacral and pudendal nerve stimulation. They report that the IRBs of Michigan Medicine and Trinity Health Chelsea Hospital approved the study and that all necessary participant consent was obtained. Two authors have a research contract from Medtronic unrelated to funding of this work; the manuscript states that contract did not fund the present study. The article is a medRxiv preprint and has not been peer reviewed. The authors state that all data produced in the study are available on reasonable request.
This report is based on intra-operative recordings from nine participants; the authors present the sample size and p-values for key comparisons in the abstract. The study team labeled this as an initial human comparison and the work appears exploratory. The abstract does not supply a complete set of stimulation parameter tables or detailed participant demographics—those items were not reported in the abstract excerpt and may be in the full manuscript or supplementary files. The preprint status means that independent peer review is pending. Open questions that remain, as implied in the report, include whether the recruitment-efficiency advantage of SNS translates into differential clinical outcomes for bladder or bowel dysfunction, how reproducible the intra-operative measures are across a larger, more diverse cohort, and whether lead design or positioning strategies can be optimized to equalize or augment recruitment efficiency for PNS. These specific clinical outcome and optimization questions were not answered in the source and remain for future work.